DSA8300 Digital Serial Analyzer and Modules Service Manual
10. Measure the TDR Clock Out frequency. (Recheck)
11. Verify that the 10 MHz measurement is within 2 kHz.
12. Save the mainframe factory calibration constants by entering the following
GPIB command:
"CAL:SAV:FACTORY:MAIN"
13. Update the mainframe calibration time/date/temperature stamp by entering
the following GPIB command:
14. Enter the following GPIB command to turn the instrument cal constant
protection on:
SYST:PROT ON
End of Procedure
Adjustment Procedures
"CAL"UPDATEINFO:MAIN"
3–5
Need help?
Do you have a question about the DSA8300 and is the answer not in the manual?