Firmware Updates; After Repair; Table 4-4: Action Required For Assembly Replaced - Tektronix DSA8300 Service Manual

Digital serial analyzer and modules
Hide thumbs Also See for DSA8300:
Table of Contents

Advertisement

Troubleshooting

Firmware Updates

After Repair

4–58
To update the firmware, install the firmware disc in your instrument and follow
the displayed instructions.
To order a firmware update, contact your Tektronix service center.
After removal and replacement of an assembly due to electrical failure, you must
perform an adjustment or software update as indicated in the following table.

Table 4-4: Action required for assembly replaced

Action required for assembly replaced
Assembly replaced
Front panel assembly
Time Base board
Optical Front-End board
Electrical Front-End board
Power PC board
ATX processor assembly
Display panel or display
system
Low Voltage Power supply
Fans
DSA8300 Digital Serial Analyzer and Modules Service Manual
Adjustment required
No
Yes
Yes
Yes
Yes
No
No
No
No
Software update required
None
None
None
None
Boot ROM and authorization
code
Windows, instrument
application
None
None
None