ICP DAS USA I-8014W User Manual page 30

250 ks/s, 16-bit, 8-channel differential/16-channel single-ended analog input module
Table of Contents

Advertisement

Step5. Select the I-8014W slot number and input range from the drop-down list
located in the upper part of the window.
Step6. Determine two values (points) in the range of the selected input type for
calibration process.
For example, after selecting 0 (-10V ~ +10V), we would like to use +8V and
-8V as the two calibration points:
Step7. Make calibration source output one of the two points (ex. 8V)
I-8014W API User Manual, v 1.0.1, September 2011
Copyright © 2011 ICP DAS Co., Ltd. All Rights Reserved.
E-mail: service@icpdas.com