ICP DAS USA I-8014W User Manual page 19

250 ks/s, 16-bit, 8-channel differential/16-channel single-ended analog input module
Table of Contents

Advertisement

Step 2. Launch MiniOS7 Utility on PC, download the calibration program to the
control unit and run it.
The MiniOS7 Utility could be downloaded from the following web site. Select the
proper calibration program for your controller.
● MiniOS7 Utility:
● 8014cal.exe: the calibration program for I-8000 unit, located at the same folder
as I-8014W demo programs. (See
● iP_8014cal.exe: the calibration program for iP-8000 unit, located at the same
folder as I-8014W demo programs. (See
a. Launch MiniOS7 Utility on
PC, choose connection →
New connection.
b. Select the COM port on PC connected to the control unit from the drop-down
list, configure the communication parameters as follows, and click OK.
I-8014W API User Manual, v 1.0.1, September 2011
Copyright © 2011 ICP DAS Co., Ltd. All Rights Reserved.
http://www.icpdas.com/download/minios7.htm
Demo Programs Location, page
Demo Programs Location, page
E-mail: service@icpdas.com
12)
12)
19