e. Make the calibration source output the other point - 8V.
f. Enter the value read by meter at the input 2nd voltage prompt and press Enter
The console displays new gain and offset values for this calibration as:
New Gain= 3xxxx, Offset= nnn, Save to EEPROM? (y/n):
g. Enter y and press Enter to accept the values and save to EEPROM
The calibration for -10V ~ +10V is complete.