b. Determine two values (points) in the range of the selected input type for
calibration process.
For example, after selecting 0 (-10V ~ +10V), we would like to use +8V and
-8V as the two calibration points.
c. Make calibration
source output one of
the two points (ex.
8V)
d. Enter the value read
by meter at the input
1st voltage prompt in
the console and
press Enter.