The procedures to set the menu item are as follows.
1. Press [Shift]+[Display] (System) to enter into the system menu interface.
2. Use left and right keys or rotate the knob to select Ext IO and press [Enter]
3. Rotate the knob to set and press [Enter] key to confirm.
4.2.5 Restore the Factory Default Settings
This menu item is used to restore the factory default settings.
The procedures to set the menu item are as follows.
1. Press [Shift]+[Display] (System) to enter into the system menu interface.
2. Use left and right keys or rotate the knob to select System Reset and press
3. Use left and right keys or rotate the knob to select the value from the follow-
4. After finishing the setting, press [Enter] key to confirm.
The detailed default parameters are listed in the following table.
On Sync Out: As the indication signal of the instrument output status. When
the output is OFF, the pins output 0V low level; when the output is ON, the
pins output 5V high level.
On Sync In: Used to control the output state of the instrument. Input 0V low
level to the pins, the instrument output is OFF; input 5V high level to the pins,
the instrument output is ON.
key to make the parameter in modification.
[Enter] key to make the parameter in modification.
ing options.
•
No: default value, indicates the power supply cancels this menu setting.
•
Yes: indicates the power supply restores the factory default settings.
No.
Parameter
The setting values on the front panel
Voltage
1
2
Current
1
Mode
2
V-Rise Time
3
V-Fall Time
4
I-Rise Time
Copyright © Itech Electronic Co., Ltd.
Default Value
0
MAX
Config menu
CV Priority
0.001 S
0.001 S
0.001 S
Functions and Features
51