Component
Task
Seat
adjustment
functions
Side
panel
and
Check that mounting screws are fastened properly
arm support
Check that screw connections between the arm support
and the control panel are tight
Check arm support for damage
Gas compression
Visually inspect the piston rod for scratches and oil
spring (option) or
leaks
actuator
Product
Check the legibility and completeness of all labels and
markings on the product
9.2 Repair
WARNING
Prohibited repairs
Severe user injuries, damage to the product due to adjustment and installation errors
► Only carry out the repairs described in this section. All other repairs may only be carried out by the qualified
personnel.
9.2.1 Replacing defective lights
NOTICE
Use of incorrect bulbs
Damage/melting of the connecting cables due to excess current flow
► Note that the LED lighting may only be replaced with original lighting.
The LED lighting is maintenance-free. If repairs are required, the qualified personnel who fitted or delivered the
wheelchair can help.
9.2.2 Replacing the battery
Batteries may only be replaced by qualified personnel.
9.3 Troubleshooting
INFORMATION
In the event of communication problems in the bus system of the controls, the system triggers an emergency stop
and thus prevents any uncontrolled functions.
► Note that after every emergency stop, you have to turn the power wheelchair control unit on again.
► If the driving function is still not available after turning the control unit on again, activate pushing mode by
releasing the brake.
► Inform the qualified personnel immediately.
Faults are indicated on the control panel LCD screen, on the LCD monitor (if installed) or on the attendant control
(if installed). The following table shows the individual notifications as well as the associated fault sources and pos
sible causes and measures.
If the measures described here do not resolve the faults completely, contact your specialist dealer. The specialist
dealer is able to read the exact error codes with a handheld programming device and can perform a targeted sys
tem analysis.
The control unit stores a list of all faults that occur. The specialist dealer reads this information, for example during
a general overhaul of the power wheelchair. Based on the saved data, the specialist dealer determines future ser
vice and maintenance intervals.
9.3.1 Types of notifications
Warning
A warning indicates a status or malfunction of one or several components of the power wheelchair. The function of
components without errors is not restricted.
C1000 DS
Maintenance and repair
Prior to
Weekly
every use
X
X
Monthly
X
X
X
71