Display symbol
LCD monitor
display
9.3.4 Attendant control error overview
Battery LED
LED
Red is flashing
Green is flashing
Mode LED
LED
Red is flashing
Seat function LEDs
LED
Backrest LED is flashing
C1000 DS
Error/warning
Back angle adjustment
motor fault
Seat tilt motor fault
Seat height adjustment
motor fault
Power
footrest
fault
Drive motor fault
Brake fault
Emergency stop
Error/warning
Battery undervoltage
Battery overvoltage
Error/warning
Various product faults
Error/warning
Back
angle
adjustment
motor fault
Cause
Faulty cabling or plug
connections
Defective actuator
Faulty cabling or plug
connections
Defective actuator
Faulty cabling or plug
connections
Defective actuator
motor
Faulty cabling or plug
connections
Defective actuator
Faulty cabling or plug
connections
Defective drive motor
Open brake release
Defective brake
Severe controller/hand-
held
control
and/or drive motor mal
function
Cause
Battery deep discharge
Voltage
too
high
complete
charging
driving downhill)
Cause
More detailed information about the existing fault is
shown on the LCD screen on the control panel (see
Page 72 f.).
Cause
Faulty cabling or plug con
tact
Maintenance and repair
Possible measure
Check cabling/plug con
nections
Contact specialist deal
er
Check cabling/plug con
nections
Contact specialist deal
er
Check cabling/plug con
nections
Contact specialist deal
er
Check cabling/plug con
nections
Contact specialist deal
er
Check cabling/plug con
nections
Contact specialist deal
er
Close brake release
Contact specialist deal
er
Check cabling/plug con
device
nections
Contact specialist deal
er
Possible measure
Charge as soon as pos
sible
(after
Continue driving slowly
and
Possible measure
Possible measure
Check cabling/plug con
tacts
73