Maintenance and repair

Straps with metal closures may not be washed in the washing machine as the penetration of water could
cause corrosion and subsequent malfunctions.
Clean the belt straps by gently dabbing them with warm soapy water (with some disinfectant) or carefully wipe
with a dry, clean, absorbent cloth.
Cleaning a belt system with plastic closure
Depending on the model, belts/straps with plastic buckles can be washed in the washing machine between
40 °C and 60 °C.
Recommendation: Use a laundry bag or net and mild detergent.
INFORMATION
Alternatively, the belt straps can be cleaned by gently dabbing them with warm soapy water (with some disinfect­
ant) or carefully wiped with a dry, clean, absorbent cloth.
Additional cleaning instructions
Allow the belts to air dry. Ensure that the belts and pads are completely dry before installation.
Do not expose the belts to direct heat (e.g. sunshine, stove or radiator).
Do not iron or bleach the belts.

9.16.2 Disinfection

1) Thoroughly clean the pads and handles before disinfecting.
2) Wipe all parts of the wheelchair with a disinfectant.
Important information about disinfecting
If the product is used by more than one person, using a conventional disinfectant is required.
Only use colourless water-based disinfectants. Observe the instructions for use provided by the manufacturer.
10 Maintenance and repair

10.1 Maintenance

WARNING
Insufficient maintenance
Severe user injuries, damage to the product due to failure to observe maintenance intervals
► Only carry out the maintenance tasks described in this section. All other maintenance and service tasks may
only be carried out by qualified personnel.
► The functionality and operating safety of the product must be verified and a service performed at least
once per year.
► For users with a changing anatomy (for example body dimensions, weight) or users with a changing clinical
picture, have the product inspected, adjusted and serviced at least once every six months.
The function of the product should be checked before each use.
The product may not be used if defects are noted. This applies in particular in case of instability of the product
or altered driving characteristics as well as problems with the user's seating position or the stability of the seat.
Inform the qualified personnel promptly for the rectification of defects.
This also applies if loose, worn, bent or damaged components, cracks in the frame or broken frame compon­
ents are identified.
Some maintenance tasks can be carried out to a certain extent by the user at home (see the sections "Mainten­
ance intervals" and "Maintenance tasks").
Failure to maintain the product can lead to injuries for the user of the product.

10.1.1 Maintenance intervals

The functions described below must be checked by the user or an attendant at the specified intervals:
Inspection task
Functional test of the wheel locks
Sagging of the seat or back support upholstery
Stability of the leg supports
30
Before each use
X
Monthly
Quarterly
X
X
X
Centro S1