Service Manual
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
9.9
Replacing the Gasket
9.10
Replacing the Battery Packs
Danger of serious injury from electric shock!
•
Prior to performing any work on the light, disconnect it from the on-site power supply and make sure
that it cannot be accidentally switched on while working on the light.
•
Prior to performing any work on the light head, be sure to remove the sensoGrip after disconnecting
the light from the on-site power supply.
Danger of burns!
•
When working on the battery pack, especially when making measurements, be sure to avoid short-
circuiting the battery poles (RD = +, BK = -).
•
Therefore, take off your watch, jewelry, etc., before starting any work on the batteries and use your
tools (screwdriver, etc.) with care.
Revision 2
•
Lift gasket (2) off the bottom shell (1) of the
housing.
•
Install new gasket (2), making sure that it
attaches well to the bottom shell (1).
•
If required, provide the gasket with cutouts (can
be done with end-cutting pliers used in
electronics). This may be necessary for the
parallel-grooved pins of the bearing block (1) in
particular.
81