Can Bus System/Rs232; Cat Wiring From Hw 30 - KLS Martin group marLED E3 Service Manual

Operating lights
Table of Contents

Advertisement

Service Manual
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
6

CAN Bus System/RS232

6.1

CAT Wiring from HW 30

The CAT system for the CAN bus can be recognized by 1–2 RJ45 cables projecting from the ceiling tube to
be mounted. One of these cables is sufficient for wiring up to 2 devices.
These cables must be connected to the matching couplers extending from the central bearing shaft.
If a camera is part of the configuration (either in the light head or with its own arm), then its CAN bus is
integrated into the CAT system. The video signals are transmitted by separate cables (3).
CAN bus wiring is done at this point with 1 RJ45 cable for 1–2 devices or with 2 RJ45 cables for 3–4 devices.
The cables and couplers are lettered to prevent mix-up.
Revision 2
Ceiling tube for 2 lights and 1 camera
1 RJ45 cable (1–2 items)
2 Power supply for operating lights
3 Camera video signals
Mounting the central bearing shaft (5) on the ceiling tube
Plug the RJ45 cables (1) and (2) into the matching (identically
lettered) couplers (6) and (7) extending from the bearing shaft.
4 Input 2: plug
5 Input 1: plug
6 Input 1: coupler
7 Input 2: coupler
8 Central bearing shaft (central axis)
Make sure that the terminating resistor is configured correctly on
the CAN board,
see section 6.1.5 "Standard CAN Board, no Galvanic Isolation",
page 32
or
section 6.1.8 "CAN Board with Galvanic Isolation", page 34.