Service Manual
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
7
Central Bearing Shaft/Tracking Arm/Spring Arm Wiring
7.1
Central Bearing Shaft (Slip Rings)
7.1.1
Replacing the Contact Block
The contact block must be replaced in the event of:
•
tracking arm replacement
•
system conversion
•
slip-ring connection wire breakage
•
slip ring replacement
Installing the contact block using the mounting tool supplied!
•
Make sure that the orange tongues (1) of the contact carrier are well positioned on the corresponding
yellow slip rings (3).
•
Use the mounting tool (2, REF 08-666-26-04).
Fig. 7-1:
Contact block on tracking arm
Revision 2
1 Contact tongues on contact carrier
2 Mounting tool
3 Slip rings
4 Cross-recessed head screw for
contact block fixation
•
Reassemble the contact block
and the spring arm one-to-one.
53