Designation
Illumination depth, L1+L2
at 20%
at 60%
Protection class
Ambient conditions
Storage temperature
Operating temperature
Air pressure
Max. operating height
Conformity to standards
IEC 62471:2006 (photobiological safety)
All values were determined in accordance with IEC 60601-2-41.
Subject to technical and design changes.
Photometric values are subject to a tolerance of ±10%.
*
In accordance with IEC 60601-2-41, the maximum illuminance must not exceed 160,000 lx.
Here, the tolerance is therefore -10%.
** Shadow dilution relative to Ec max.
184
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E9
up to HW 39
330+500 mm
180+230 mm
I acc. to IEC 60601-1
-10°C to +50°C (+14°F to +122°F) at 20–90% humidity
(non-condensing)
+5°C to +40°C (+41°F to +104°F) at 30–80% humidity
(non-condensing)
4,000 m above sea level
IEC 60601-1:2005 (Third Edition)
+ CORR. 1:2006 + CORR. 2:2007 + A1:2012
IEC 60601-1-2:2007, modified
IEC 60601-2-41:2009+A1:2013
according to risk group 1
marLED® E3 on ceiling tube Ø 110 mm
marLED® E9i
up to HW 39
330+430 mm
180+200 mm
620–1,060 hPa
Service Manual
marLED® E9i
from HW 40
370 + 630 mm
250 + 290 mm