10.2
Connecting a Wall Control Panel HW 30–HW 36
The terminating resistor of the wall control panel should always be switched on:
At least one of the two S1 DIP switches must be set to ON.
100
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
1 Wall control panel, rear view
2 Adapter cable with RJ45 coupler
3 RJ45 cable from CAN board
The wall control panel (1) HW 30–HW 39 can be
retrofitted with an RJ45 adapter (2).
To do this, connect the adapter cable (2) with
wire-end sleeves to the X2 terminal block of
the COM interface.
1 Wall control panel, opened
2 RJ45 adapter
S1 Terminating resistor (off = both off)
X2 Terminal block:
1 = CAN H (WH)
2 = CAN L (BN)
3 = VDC+ (RD)
4 = VDC- (BU)
Connecting the wall control panel to the light
system
Plug the RJ45 cable from the socket "wall
control panel" of the CAN board into the RJ45
adapter (2).
Service Manual
Revision 2