Variants
Date
October
all
2012
all with wall
November
control panel,
2012
COM interface
December
all
2012
172
SW release
Description of change
marLED® COM interface software update from SW 0270 to
SW 0306
Activation of camera control function of wall control panel.
This software version also includes the marLED® V communication
SW 05
software, so there will be only one common marLED® COM
interface software in the future.
surgiCam daughterboard software update from SW 0028 to
SW 0029
Problem of irregular camera zoom speed eliminated.
marLED® COM interface software update from SW 0306 to
SW 0310
SW 06
This software version also includes the marLED® V communication
software, so there will be only one common marLED® COM
interface software in the future.
Control electronics software update from SW 0037 to SW 0038
In this version, the internal pull-up resistors for the switching signals
SW 07
were deactivated. This eliminates the problem of unwanted light
adjustments that occurred automatically during operation, without
any sensoGrip action.
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
Service Manual