Risk of material damage in the event of light failure!
The use of non-identical battery packs can lead to light failure or shorten the service life of the light.
•
Always replace the battery packs as a set (2 items).
Danger of injury due to short circuit! Explosion hazard!
•
Be sure to use only genuine Gebrüder Martin battery packs!
Non-compliance will void any warranty claims against Gebrüder Martin.
Returning used battery packs!
•
Batteries whose capacity has dropped below 80% of their rated capacity (for more information, see
marLED® battery, Instructions for Use) must be marked "To be checked by supplier" before returning
them.
•
Open contacts must be protected with masking tape.
•
Use suitable transport packaging. If undamaged, the original packaging can be used for this purpose.
•
Pack tightly so that the batteries cannot move around in transit.
•
Remove the sensoGrip.
•
Open the light housing, see section 9.8 "Opening the Light Housing", page 79.
This includes disconnecting the battery cables, see section 9.5 "Disconnecting the Battery Cable
(marLED® E battery)", page 77.
82
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
Service Manual
Revision 2