Service Manual
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
9.5
Disconnecting the Battery Cable (marLED® E battery)
9.6
Connecting the Battery Cable (marLED® E battery)
Revision 2
•
Press safety catch (2) of the PCB plug (1) to
unlock it.
•
Withdraw plug (1) from the control electronics
board as indicated by the arrow.
This isolates the battery pack from the light.
If required, the battery pack can now be replaced.
nd
•
Handle the 2
battery pack in the same way.
•
Be aware of the anti-rotation element (2) of
the PCB plug (1).
•
Push PCB plug (1) in place on the control
electronics board as indicated by the arrow.
Verify proper engagement of the safety catch by
carefully pulling at the connecting cable.
nd
•
Handle the 2
battery pack in the same way.
77