17.5
Technical Data for marLED® E3
Designation
Power supply
Power consumption on 220–240 VAC
Power consumption on 100–130 VAC
Light sources
Anticipated LED service life
Max. illuminance Ec at a distance of
1,000 mm
Total irradiance at maximum illuminance
Luminous efficacy
Light field diameter D10
Light field diameter D50
Color temperature
Color rendering index Ra and R9
Dimmability
Operation
Illumination depth, L1+L2
at 20% of Ec
at 60% of Ec
Protection class
Sterilizable handle option
Disposable handle option
Ambient conditions
Temperature for storage and
transportation
Operating temperature
Air pressure
Max. operating height
Conformity to standards
IEC 62471:2006 (photobiological safety)
All values were determined in accordance with IEC 60601-2-41.
Subject to technical and design changes.
Photometric values are subject to a measuring tolerance of ±10%.
Revision 2
marLED® E3
100–240 VAC, 50–60 Hz
32 VA
32 VA
High-performance LED, arranged into 3 Light Engines
60,000 lx
ca. 185 W/m²
330 lm/W
150 mm
-
4,500 K
Non-sterile via control panel: ON/OFF,
-
-10°C to +50°C (+14°F to +122°F) at 20–90% humidity (non-
+5°C to +40°C (+41°F to +104°F) at 30–80% humidity (non-
4,000 m above sea level
IEC 60601-1:2005 (Third Edition) + CORR.1 (2006) + CORR.2 (2007) +
A1:2012 + IEC 60601-1-2:2007
IEC 60601-2-41:2009
according to risk group 1
marLED® E3
60,000 h
100,000 lx
ca. 364 W/m²
275 lm/W
95
30–100%
Dimming 30–100%
825 + 820 mm
600 + 320 mm
I acc. to IEC 60601-1
yes
yes
condensing)
condensing)
620–1,060 hPa
39 VA
32 VA
170 mm
105 mm
4,300 K
191