Once the software update has been finished successfully, click OK to close the window.
•
Change the switch position Program (5) on the programming board to Run (7), see "Settings on
programming adapter", page 127.
•
After programming, press the Reset button 2x.
•
Unplug the programming adapter and operate the power ON/OFF button of the light.
•
Turn on the light again to perform function test.
It is absolutely essential to press the Reset button 2x after programming.
130