SW6: Configuration marLED®
DIP
OFF
3
Reserve, currently ineffective
4
No level-selection boost
5
Light field size setting via sensoGrip (presetting)
7
without surgiCam daughterboard
Current DIP switch assignment from HW 40
DIP 6
DIP 8
OFF
OFF
ON
OFF
OFF
ON
ON
ON
6.4.3
DIP switch SW1 marLED® for Terminating Resistor
DIP 1
DIP 2
OFF
OFF
ON
OFF
OFF
ON
ON
ON
To switch on the terminating resistor, see
•
section 6.1.6 "Terminating Resistor S1 without Coupler", page 32
•
section 6.1.9 "Terminating Resistor S1 with Coupler", page 34
Restart required if settings have been changed!
If the CAN address or DIP switch settings have been changed or components have been disconnected or
connected, the entire light system must be restarted to make the changes effective:
•
Switch power off and on again.
–
The new settings are now effective.
50
Configuration E9/E9i/E15
Boost function available (E9 up to HW 39, E9i from HW 40)
No boost function available (E9i up to HW 39)
E15
Avoid – leads to malfunction!
Terminating resistor R
A
OFF – mandatory with surgiCam daughterboard
ON
ON
ON
Operating Lights marLED® E3, E9, E9i, E15, battery and
marLED® E3 on ceiling tube Ø 110 mm
ON
Reserve, currently ineffective
Level-selection boost for marLED® E9/E15
Color temperature setting via sensoGrip
surgiCam daughterboard active
Service Manual
Revision 2