Type Plate; Mains Connection; Equipotential Grounding Pin; Foot Switch Connection - KLS Martin group Laser Surgical Unit diomax 1550 Instructions For Use Manual

Table of Contents

Advertisement

Instructions for Use
Laser Surgical Unit diomax
4.4.1

Type Plate

4.4.2

Mains Connection

4.4.3

Equipotential Grounding Pin

4.4.4

Foot Switch Connection

Revision 2
®
1550
The type plate is on the rear of the device
(see section 4.4 "Rear of Device", page 24).
See also section 11 "Technical Data", page 66.
The device is equipped with a universal power supply unit.
The mains voltage range is 100 – 240 V, 50/60 Hz.
The mains connection (2) has two fuses:
2 × T6.3A H250V.
Connect the mains connection on the rear of the device
to the mains of the building using the supplied mains
connection cable.
The device can be connected to the potential equalization
in the operating room via the equipotential grounding pin
using the equipotential grounding cable (included in the
scope of delivery).
Socket for the connection of the foot switch for laser
activation. The function of the foot switch is described in
section 5.8 "Foot Switch", page 40.
Place the foot switch near the operation field in a
position that allows the user to easily reach it with his
or her foot, and insert the connection cable with the
black plug (guide groove on top of plug) into the
connection socket Foot switch.
For removal do not pull on the cable but loosen the
plug connection by pulling on the plug casing.