Surgical Laser Unit Limax
5.5.10

Line Fuses for the Optical System

5.5.11

Equipotential Bonding Pin

5.5.12

NIST Socket (AIR)

V. 1.0
®
60
If current consumption is too high in the laser circuit, the
optics line fuses (Fig. 5.5, (10)) will be tripped. Pushing
them back in again reactivates them.
The equipotential bonding pin (Fig. 5.5, (11)) is intended to
connect the Limax
of the OR using the equipotential bonding cable provided.
The NIST socket (AIR) (Fig. 5.5, (12)) serves to connect
the Limax
A special air hose (black or black / white acc. to ISO) must
be used for this purpose. It is to be connected to the
compressed-air (AIR) outlet (MEDAP, DIN AGA, BOC, ALI).
Suitable hoses are available on request.
Be sure to observe the instructions for
connecting/removing the compressed-air hose!
See section 6.6 "AIR (Purge Air)", on page 49.
®
60 to the equipotential bonding system
®
60 to the compressed-air supply of the OR.