Technical Data

Description
Laser type
Laser wavelength
Laser output
Pulse forms
Beam guidance
Laser beam quality
Laser fiber connection
Control and monitoring
Cooling
Mains connection
Mains power
Mains fuse
Power consumption
Laser class
Protection class
Degree of protection
Classification acc. to German
Medical Devices Act
Pilot laser wavelength
Pilot laser power
Pilot laser class
66
®
diomax
1550 data
Diode laser
1,550 nm ± 2%
0.1 – 15 W ± 20%
Continuous mode:
Continuous rate:
Pulsed mode:
Pulse rate:
Cycle mode:
Single mode:
Laser fibers, focusing handpiece
Numerical laser aperture ≤ 0.22
SMA-X socket, mechanically coded SMA socket
2 microprocessors
Semiconductor/air cooling
100 – 240 V ± 10%; 50/60 Hz
max. 6.3 A
2 × T 6.3A H250 V
650 W
4
I
IP X1
II b
635 nm
3 mW ± 20%, 2 – 100% adjustable, continuous and flashing at
0.5 Hz/1 Hz/2 Hz/3 Hz/4 Hz/5 Hz
3R
Laser Surgical Unit diomax
Output:
0.1 – 15 W
Output:
1 – 15 W
Energy:
10 – 150 J/cm
Pulse time:
5 ms – 10 s
Pulse pause:
5 ms – 10 s
Output:
1 – 15 W
Pulse time:
5 ms – 10 s
Pulse pause:
5 ms – 10 s
Energy:
10 – 150 J/cm
Pulse time:
5 ms – 10 s
Pulse pause:
5 ms – 10 s
Cycle time:
20 ms – 40 s
Cycle pause:
5 ms – 10 s
Output:
1 – 15 W
Pulse time:
5 ms – 10 s
Instructions for Use
®