9. Stylus Tip Replacement
1. Lift the Stylus Tip up.
2. Remove two locking screws from the Stylus Tip.
3. Replace the Stylus Tip with a new one.
4. Assemble in reverse order.
Caution
None
Adjustment needed
Perform Pattern & Frame Calibration.