4. R-axis Rotation Cable Replacement
1. Remove the Bottom Plate.
2. Disconnect the R-axis Rotation Cable connector linked between DC motor I/F Board and Middle I/F
Board.
3. Remove the R-axis Rotation Cable.
Be sure to remember the tied shape of the R-axis Rotation Cable.
4. Replace it with a new one.
5. Tie the R-axis Rotation Cable with cable tie in the same way of the original shape. (Important)
6. Be sure to check whether the cable is free enough when it rotates fully and if there's any bump with
other parts.
7. Assemble the bottom plate in reverse order.
Caution
Step 6 is extremely important for the cable lifetime.