Setting The Temperature Output Terminal - Yamato INE800 Manual

Eco incubator
Table of Contents

Advertisement

The temperature output terminal has a function to output the temperature displayed on the controller to
the recorder.
1 Turning power on
2 Setting temperature output
Example
Output:Current of 4~20mA will be output for temperature indication of -5~65℃ when PV,
4-20.H:65 and 4-20.L:-5.
Measuredtemp.
When you want to connect a terminal to the voltage input of the recorder, install a fixed resistor (shunt
resistor) of 600Ω or lower.
Example When a shunt resistor of 250Ω is employed:
Indication Output current Voltage input to recorder
-5℃
65℃
Turn the ELB at the upper right side of the unit "On".
Keeping the
Main display shows the internal temperature and the Sub
display shows the set temperature.
① Press the
② Press the
③ Press the
④ Press the
⑤ Press the
65℃
-5℃
4mA
4mA
20mA
4. Operating procedures

Setting the temperature output terminal

key pressed longer will turn power ON.
key to indicate Sub display [toUt] in the
Sub display and decide on it with the
key to select an item to output.
:Outputs the set temperature.
:Outputs the indicated temperature.
:Outputs nothing.
:Outputs heater operation amount *※
key, enter the upper limit (initial
value:65℃) for the output temperature and then decide
on it with the
key.
* Enter 100% for the output of heater operation amount
key, enter the lower limit (initial
value:-5℃) for the output temperature and then decide
on it with the
key.
* Enter 0% for the output of heater operation amount
key to return to the initial screen.
Esc
1V
5V
35
Current output
20mA
key.