When the "start temperature" and the "set temperature" are the same, the temperature will be
maintained at that level until the set time has elapsed. At this time, if the indicated
temperature becomes ±1℃ or more to the set temperature, the mode will change to "wait"
and counting of the remaining time will be aborted.
If you set "0" for the set time, the unit will operate at its full power from the "start temperature"
to the "set temperature". When you set wait ON, the "wait" status will continue until the
indicated temperature come within the range of±1℃ of the set temperature.
Start
temp
The diagram below shows the repeat concept. The first execution of the repeat intervals is
not counted as repetition counts.
Step # 1
Step # 2
No repeat command
No repeat command
Temp
Start
temp
Set time
Temp
Time to reach set temp
During wait
Step # 3
No repeat command
Repeat section repeated
Five times
29
4. Operating procedures
How to input a program
Set temp
Set temp
Step # 4
Repeat start
Step :2
Times:5
Time
Time
Step # 5
No repeat command