Setting Auto Defrost - Yamato INE800 Manual

Eco incubator
Table of Contents

Advertisement

1 Setting auto defrost

4. Operating procedures
※You cannot change the current auto defrost setting while
the unit is in operation. First, stop operation before setting.
① Press the
key several times until [dEFCL] appears in
the Sub display and then press the
② When [A-dEF] appears in the Sub display, press
the
key and select an auto defrost mode from three
patterns.
non : auto defrost OFF
tIME: cycle defrost (Required settings-dEF:Cy, dEF:tM)
CLK : Clock defrost(Required settings-dFS.t、dEF:tM)
dFS.t : Defrost time(initial value: 00:00)[TIME]
dEF:Cy: Ordinary run time (initial value:24hrs)[TIME]
Counted at the start of operation
dEF:tM: Defrost time(initial value:5min)[MIN]※
Example 1 Cycle defrost
A-dEF :t|ME
Every 12 hours after start of operation.
dEF:Cy:12 ⇒ Defrosting is performed for five minutes.
dEF:tM:5
Example 2 Clock defrost
A-dEF :CLK
Defrosting is performed for five minutes
once every day at 17:00.
dFS:t :17:00 ⇒
dEF:tM:5
※ When setting to a larger value, select a value taking
possible
increase
consideration. The margin of increase is about3 ℃ for
five-minute defrosting, which differs depending on the
internal set temperature or other conditions.
③ After completing setting, press the Esc key twice to return
to the initial screen.
16