Time-up output is a function that outputs a signal when operation is terminated (END indication).
Connecting terminals
External alarm is a function that outputs a signal when an error occurs to the main unit.
Connecting terminals
4. Operating procedures

Setting time-up output (optional)

Connect the time-up output terminal at the upper
right side of the unit.
COM
Connect the a contact(relay contact).
NO
COM
Connect the b contact(relay contact).
NC
Contact capacity: AC250V 1A(resistance load)
DC30V
*About the time-up output timing
Sub display shows "END" at the end of program
run or auto stop run. A time-up signal will be
output while this "END" is displayed.
Pressing the
START/STOP
indication when time-up output is cancelled as
well.

Setting external alarm output (optional)

Connect to the external alarm output terminal at
the upper right side of the main unit.
COM
Connect the a contact(relay contact).
NO
COM
Connect the b contact (relay contact).
NC
Contact capacity:AC250V 1A(resistance load)
DC30V
* About the external alarm output
An error code will be displayed when an error
occurs. An external alarm signal is output while
the error code is displayed.
57
1A(resistance load)
key deletes "END"