Yamato INE800 Manual page 45

Eco incubator
Table of Contents

Advertisement

4.6
Details of function codes
Replies by each of function codes are shown below.
4.6.1
Readout of analogue settings
[Function code:03(03H)]
Readout will be made for "Analogue settings (2 byte:16 bit)data" whose numbers are continuous
for the specified number of values starting with the specified number. Data is divided into high 8
bits and low 8 bits, which then will be arranged in the order of their numbers to form data for the
reply message.
Example: Readout of fixed temperature operation SV temperature setting, fixed value auto start SV
temperature setting, fixed value auto start time/clock time select setting of slave 1.
(Readout of three analogue settings of slave 1 from reference #40001 to 40003)
Reference #
(relative numbers)
Data
〈RTU mode〉
Master→device
Slave address
Function code
Starting
number(H)
Starting
number (L)
Quantity (H)
Quantity (L)
CRC (L)
CRC(H)
〈Error check in the ASCII mode〉
LRC
Note: Starting number (relative number) is "Reference #-40001".
Note: The number of data pieces is the number of bytes of the data.
(This is different from the requested number. In the example, the requested number is three
while the number of pieces of the data is six.)
40001
40002
(0000H)
(0001H)
650
-50
(028AH)
(FFCEH)
Device→master (normal)
01H
Slave address
03H
Function code
00H
Number of data pieces
00H
Fixed valueSV (H)
00H
Fixed valueSV (L)
03H
Fixed value AT STA SV (H) FFH
05H
Fixed value AT STA SV (L)
CBH
Fixed value timeselect (H)
Fixed value timeselect (L)
CRC (L)
CRC (H)
F9H
LRC
41
4. Operating procedures
Setting external communication
40003
(0002H)
0
(0000H)
01H
03H
06H
02H
8AH
CEH
00H
00H
E8H
A3H
9DH