2.Data transmission system
Com system:Half-duplex asynchronous system (Boring selecting system)
Com. distance:Max. 500m (depending on influences from the environment)
Connection method:Multidrop system (Max 1 :31 stations)
Start bit:1 bit
Transmission code: ASCII (ASCII mode)
Binary (RTU mode)
Error check: LRC (ASCII mode)
CRC-16 (RTU mode)
3.Frame configuration
The RTU mode consists of a message portion only.
The ASCII mode consists of a starting character ":(colon, 3AH)", a message and a terminating
character "CR (carriage return, ODH) + LF(line feed), OAH".
RTU mode:
ASCII mode:
4.Message configuration
The configuration of an MODBUS message is as follows for both RTU & ASCII modes:
Slave address
Function code
Data
Error check
Message
: | Message
| CR
4. Operating procedures
Setting external communication
| LF
38