I.E. Environmental Requirements
Power Requirements:
Temperature:
Humidity:
The Maxwell
instrument on a clean, level surface. To avoid shortening the expected lifespan of the instrument, install
in a location that meets the following criteria:
Locate on a sturdy, level surface.
Avoid dusty areas.
Choose a location that has good air circulation and is not exposed to direct sunlight.
Avoid noisy electrical power sources (e.g., power generators).
Do not install in a location where there is large temperature variability or high humidity.
Do not position the instrument so that it is difficult to unplug from the power source.
Do not place next to heat sources.
Do not use near flammable gases or liquids.
Do not place near other electrically sensitive instruments.
II.

Hardware Overview

Maxwell
Display panel
and keypad
Door
handle
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Part# TM274
Page 8
100–240VAC, 50–60Hz, 2.1A
5–40°C
up to 80% relative humidity
16 Instrument is intended for indoor use only. Wipe up spills immediately. Install the
®
®
16 Instrument Front
Maxwell
®
16 Instrument Back
Power
connection
Power
switch
Label
Serial port. Connect to
computer using the RS-232
Cable to obtain firmware upgrades.
Printed in USA.
Revised 6/08