Note
Set the precision pressure reducer to the pressure specified on the enclosed injector data sheet
•
(see figure 5).
Note
The following table is an extract from an injector data sheet.
Injektortyp
Injektor-Nr.
I
Injector type
Injector No.
Betriebsdruck
Durchfluss
Operating
pressure
[bar]
2.4
2.6
2.8
3.0
3.2
3.4
3.6
3.8
4.0
Überprüfung des Verdünnungsfaktors
Check of the dilution ratio
Kritische Düse
Verd.gas
Critical nozzle
Dilution gas
5.1 l/h
100 % N
Figure 7
Extract from an injector data sheet
28
The total heating time is about 2 hours. After about 1 hour, the lower
parameter value ( 30°C [54 °F] below set value) is exceeded.
For safety reasons, the low pressure gauge must show a low pressure of
> -0.6 bar. With a low pressure of < -0.6 bar, the critical orifice will not
function properly. If the necessary low pressure will not be reached, the
dilution gas pre-pressure must be increased.
Für
Verdünnungssystem-Nr. ....../............
.....
For dilution system No.
Unterdruck ohne kritische Düse
Flow
Vacuum without critical orifice
[l/h]
415
435
460
490
510
535
560
585
605
Verdünnungsgasdruck
Dilution gas pressure
3.1 bar
2
SP190-H/DIL-EX2 T* | 1.01.00
Unterdruck mit kritischer Düse
Vacuum with critical orifice 5.1 l/h
[bar]
-0.62
-0.65
-0.68
-0.80
-0.79
-0.79
-0.78
-0.77
-0.77
Messgasdruck
Sample gas pressure
Messgas
Verdünnung
Sample
Dilution
100 % O
100:1
2
Typ
SP190-H/DIL-
EX2 T*
Type
[bar]
-0.61
-0.64
-0.67
-0.77
-0.77
-0.77
-0.76
-0.76
-0.75
atmosphärisch
atmospheric
Messwert d. verd. Gases
Meas. value of the dil. gas
1.0 % O
2
www.mc-techgroup.com