Figure 7 Extract From An Injector Data Sheet - M&C SP Series Instruction Manual

Gas sample probe
Hide thumbs Also See for SP Series:
Table of Contents

Advertisement

Note
Set the precision pressure reducer to the pressure specified on the enclosed injector data sheet
(see figure 5).
Note
The following table is an extract from an injector data sheet.
Injektortyp
Injektor-Nr.
I
Injector type
Injector No.
Betriebsdruck
Durchfluss
Operating
pressure
[bar]
2.4
2.6
2.8
3.0
3.2
3.4
3.6
3.8
4.0
Überprüfung des Verdünnungsfaktors
Check of the dilution ratio
Kritische Düse
Verd.gas
Critical nozzle
Dilution gas
5.1 l/h
100 % N
Figure 7
Extract from an injector data sheet
28
The total heating time is about 2 hours. After about 1 hour, the lower
parameter value ( 30°C [54 °F] below set value) is exceeded.
For safety reasons, the low pressure gauge must show a low pressure of
> -0.6 bar. With a low pressure of < -0.6 bar, the critical orifice will not
function properly. If the necessary low pressure will not be reached, the
dilution gas pre-pressure must be increased.
Für
Verdünnungssystem-Nr. ....../............
.....
For dilution system No.
Unterdruck ohne kritische Düse
Flow
Vacuum without critical orifice
[l/h]
415
435
460
490
510
535
560
585
605
Verdünnungsgasdruck
Dilution gas pressure
3.1 bar
2
SP190-H/DIL-EX2 T* | 1.01.00
Unterdruck mit kritischer Düse
Vacuum with critical orifice 5.1 l/h
[bar]
-0.62
-0.65
-0.68
-0.80
-0.79
-0.79
-0.78
-0.77
-0.77
Messgasdruck
Sample gas pressure
Messgas
Verdünnung
Sample
Dilution
100 % O
100:1
2
Typ
SP190-H/DIL-
EX2 T*
Type
[bar]
-0.61
-0.64
-0.67
-0.77
-0.77
-0.77
-0.76
-0.76
-0.75
atmosphärisch
atmospheric
Messwert d. verd. Gases
Meas. value of the dil. gas
1.0 % O
2
www.mc-techgroup.com