15.3 Connecting the test gas line
To connect the test gas line, a threaded tube connector with Ø 6 x 1 mm (DN4/6) is available
•
(see Figure 3 and Figure 4)
If using a PTFE tube as sample line, a metal tubing sleeve for pneumatic fittings needs to be
•
inserted at the end of the tube to prevent it of being pressed together.
•
The temperature-resistant, stainless steel connector from M&C has a double-blade ring system
to ensure reliable sealing. First finger-tight the nuts of these connectors, then use a flat spanner
to turn the nuts exactly 1¼-turns. The nuts are now correctly mounted.
www.mc-techgroup.com
Make sure that all gases or gas compositions, which enter the
sample gas probe, do not react with the environment or the
process.