Figure 9 Dilution Block And Special Tool - M&C SP Series Instruction Manual

Gas sample probe
Hide thumbs Also See for SP Series:
Table of Contents

Advertisement

Figure 9
Dilution block and special tool
Please proceed as follows:
Use the special tool, provided by M&C, to remove the critical orifice and/or the eductor from the
dilution block.
Clean the orifice and/or the eductor in an ultra sonic bath.
Check the O-rings and change them if necessary.
Push the new or cleaned critical orifice and the eductor with its nozzle seat and O-ring, carefully
back into the block by using the special tool as described in Figure 9.
Now, the dilution block can be reassembled in reverse order.
www.mc-techgroup.com
Mounting srew tool for the critical orifice or
eductor; Part.No.: 93S4440
SP190-H/DIL-EX2 T* | 1.01.00
Eductor marked with a slot to
eductor marked with a slot to differ it
differ it from the bypass eductor
from the bypass eductor