8
Technical Data
8.1
Gas Sample Probe with Weather Protection Cover
Gas Sample Probe Series SP
Part No.
Weather protection cover with
insolation
Mounting outdoors
Degree of protection
Sample temperature
Gas flow
Recommended for dust loadings
Filter chamber volume
Filter element
Probe heating temperature
Ready for operation
Temperature alarm contact, alarm
point
Temperature alarm, contact rating
Connection sample outlet diluted
Connection test gas
Junction box
Power supply
Power consumption
Electrical connection terminals
Electrical equipment standard
Mounting flange
Critical orifice type
Dilution rates
1)
Sample flow rate
2)
Possibility to adapt the dilution factor
www.mc-techgroup.com
®
SP190-H/DIL-EX2
T2
02S1960
Yes, the probe must not be operated without the weather
protection cover.
Yes
IP54, EN 60529
Max. 600 °C* [1112 °F*] The temperature of the process must
not heat the probe beyond the T-Class. Appropriate heat
decouplings must be provided.
Max. 500 Nl/h
Up to 1 g/m
* max.
3
70 ml
S-2K, filter porosity 2 μm, ceramic
160 to 180 °C
[320 to 356 °F]
self-regulating
after 2 hours
< 130 °C [< 266 °F],
NO
250 V-3 A AC, 30 V-3 A DC
8 mm (DN 6/8) Swagelok tube connector
6 mm (DN 4/6) Swagelok tube connector with blank plug
Fixed mounted junctionl box
110 V AC up to 240 V AC 50/60 Hz
Start up: 400 VA, during operation: 100 VA, (fuse 6 A)
Max. 2.5 mm
2
EN 61010, EN 60335-1
DN65 PN6, form B, stainless steel 316Ti with G3/4"i-
connection
a
b
500:1
200:1
1.4
2.7
With dilution gas pressure-adjustment -5 % to +30 %
SP190-H/DIL-EX2 T* | 1.01.00
SP190-H/DIL-EX2
T3
02S1965
130 to 160 °C
[266 to 320 °F]
self-regulating
< 130 °C [< 266 °F],
NO
, cable fittings 1 x M20 and 1 x M16
c
d
100:1
50:1
5.5
11
SP190-H/DIL-EX2
T4
02S1970
90 to 120 °C
[194 to 248 °F]
self-regulating
< 90 °C [< 194 °F],
NO
e
f
g
30:1
20:1
10:1
19*
28
55
3)
11