M&C SP190-H/DIL-EX2 T* probes are designed for stationary use. The proper selection of the right
sample probe for the application and the correct installation garantees a long service life and minimum
maintenance work. The ideal position for mounting the sample probe is horizontally with a tilt angle of
10 degrees towards the process.
Screw the sample probe tube directly into the ¾" inner thread of the probe flange with the ¾" flat
•
gasket, and tighten it.
Mounting the SP190-H/DIL-EX2 T* with weather protection cover: to mount the probe at the
•
sampling flange, please loosen the metal clamp around the heat insulating protection cover and
remove the cover.
If the probe connection is not the same size as the standard flange connection DN65 PN6, an
•
adapter flange (optional) is needed to be mounted to the probe. Place the flange gasket between
the adapter flange and the probe flange.
If the heated probe tube type SP30/35 is used then the probe is to be screwed to its flange (with
•
welded threaded bolt). First insert the flange gasket between the two flanges.
•
Attach the flange gasket to the probe connection.
•
Insert sample dilution probe with mounted probe tube into the connection piece and fasten the
probe using the supplied bolts and nuts.
•
SP190-H/DIL-EX2 T* with weather protection cover: after mounting of the probe at the sampling
flange put the heat insulating protection cover over the probe flange again and secure it with the
metal clamp.
Note
Figure 5
Borehole diagram of the terminal box (version without weather protection cover)
www.mc-techgroup.com
For the preferred mounting position of the probe, the sample
gas outlet is pointing downwards. This is just the preferred
mounting position, it is not necessary for correct operation.
SP190-H/DIL-EX2 T* | 1.01.00