Technical Data - M&C PMA Series Instruction Manual

Oxygen – analyzer
Hide thumbs Also See for PMA Series:
Table of Contents

Advertisement

8

TECHNICAL DATA

Oxygen analyzer Series PMA
Part No.
03A2005
03A2005a
03A2001
03A2005a
Power supply
Electrical connections
Measuring ranges
External measuring range indication
Remote range control
Indication
Output signal
Response time for 90% FSD
Accuracy after calibration
Reproducibility deviation
Reproducibility deviation without
changing the measuring range**
Influence of ambient temperature
Influence of barometric pressure
Influence of sample gas flow
Sample gas inlet pressure
Sample gas outlet pressure
Flow rate of sample gas
Temperature of sample gas
Analyzer temperature
Ambient temperature
Storage temperature
Power supply
Electrical connections
www.mc-techgroup.com
Version PMA 30/D, PMA 30A/D
PMA 30/D 230 V 50 Hz,
PMA 30/D 115 V 60 Hz,
PMA 30A/D 230 V 50 Hz,
PMA 30A/D 115 V 60 Hz
230 V AC (standard) or 115 V AC available (a)*
±15 %, 40-60 Hz, 35.5 VA
Power: 3-pole mains plug with 2 m cable; signals: 9 and 25 pole plug
Selectable for 0-1, 0-3, 0-10, 0-30 and 0-100 vol% O
One potential free contact for each selected range,
Contact rating max. 48 V DC, 0.2 A
Min. contact load 5 V/1 mA
Remote range control for all 5 ranges by means of external potential free con-
tacts max. 30 V DC 3 mA DC. (requirement: selection switch in position Ext.)
PMA 30A/D: analogue meter with a scale of 0-30 and 0-100 % for each se-
lected range, digital meter, 4 1/2 digit 9 mm high for 0-100 % O
lution 0.01 vol% O
PMA 30/D : digital meter, 4 1/2 digit 18 mm high for 0-100 % O
lution 0.01 vol% O
0-20/4-20 mA* for the chosen range, non-isolated, max. load 300 Ω
Output voltage max.15 V (delivered condition).
Switchable to a load of max. 900 Ω output voltage max.30 V.
Output current limiting adjustable 20.5 mA – 22 mA. Delivered condition
21 mA. 0-1 V DC, load > 100 kΩ, for the range of 0-100 vol%
< 3 seconds at 60 Nl/h air
1 % of the span value of measuring range after calibration or  0.02 % O
pending on which value is the higher one
Deviation: analogue = < 1% of span/digital = ±0.1 vol% O
Analog output without changing measuring range** = < 0.1 % of measuring
range / digital reading = ± 0.01 vol% O
No influence up to 45 °C
The oxygen reading varies in direct proportion to changes of the barometric
pressure.
Fluctuations of the sample gas flow in a range from 10 to 60 Nl/h nitrogen
(N
) cause a change of the oxygen reading which is smaller than 0.1 vol%
2
O
2
0.01 up to 1 bar g, (PMA 30 required admission pressure for competent flow
rate, no pump inside)
Outlet of analyzer must discharge freely into atmosphere.
Min. 10 to max. 60 Nl/h, adjustable with needle valve on the flowmeter (flow-
meter is calibrated for air and has a measuring range of 7-70 Nl/h)
-10 °C up to +50 °C, dry gas
Fixed at +55 °C
-10 °C up to +45 °C
-20 °C up to +60 °C, relative humidity 0-90% RH
Internal power unit for 230 V AC standard or 115 V AC available (a)* 15 %, 40-
60 Hz, 35.5 VA
Mains supply: 3-pole chassis plug with 2 m of cable; signals: 7, 9 and 25-pole
plugs
PMA30 | 1.01.03
2
2
2
, linear
2
reading, reso-
2
reading, reso-
2
de-
2
2
9