11

TECHNICAL DATA

Gas sample probe type
Part No.
Weather protection cover
Filter housing material
Sealing materials
Probe flange sealing material
Pre-filter/sample tubes
Sample pressure max.
Ambient temperature
Permissible process gas temperature Depending on the temperature class, however max. 200 °C [392 °F]
Filter chamber volume
Filter element, porosity
Sample gas outlet connection
Connection gas outlet at option RS
Mounting flange
Weight
Marking
Admissible medium temperature
Option heating type HEX4
Part No.
Mounting controller
Control
Power supply
Electrical connection
Marking
Case protection
Power
Max. temperature
Min. temperature
Ambient temperature
Low temperature alarm contact
Option heating type HEX1
Part No.
Mounting controller
www.mc-techgroup.com
SP3110V
(up to 185 °C [365 °F])
20S5615
Yes
Stainless steel 316/316Ti
Graphite, FKM
Graphite
Optional, see data sheet 2.14 and 2.17
0.5 to 6 bar abs.
-20 to +60 °C [-4 to 140 °F]
at the probe entry
120 cm
3
F-3SS150= stainless steel*, 3 µm S-2K150= ceramic**, 2 µm
1x 1/4" NPT female for max. 8 mm-tube connectors
6 mm Swagelok
®
connector
DN 65 PN 6, Form B, SS316Ti* >DN or ANSI possible**
7 kg [≈ 15.4 lb]
II 2G/2GD -20°C ≤ Ta ≤ +60°C EXAM BVS 04 ATEX H 045X
The admissible medium temperature is limited by the used materials
(< 200 °C [392 °F]) and further by the maximum admissible surface
temperatures as shown in Table 4.
HEX4-135
20S5510
In the Ex-zone 1, 2 or 21, 22
Self-regulating
115 V – 230 V 50/60 Hz
Cable gland, terminal range 7 – 12 mm, terminals max. 4 mm
II 2 G Ex em T3..T4 /
04 ATEX E 253
IP66, EN 60529
400 W
120 °C [248 °F]
90 °C [194 °F]
-20 to +60 °C [-4 to 140 °F]
< 60 °C [140 °F], 1 contact MC-
NO, 230 V 1.5 A AC, 0.5 A DC
HEX1-3
20S9037(a)
Outside the Ex-zone
SP3110../SP3110V.. | 1.00.02
SP3110
(more than 185 °C [365 °F])
20S5610
Graphite
HEX4-180
20S5520
II 2 D IP66 135°C...180°C EXAM BVS
160 °C [320 °F]
120 °C [248 °F]
< 100 °C [212 °F], 1 contact MC-
NO, 230 V 1.5 A AC, 0.5 A DC
2