Figure 8 Change Filter Element - M&C SP Series Instruction Manual

Gas sample probe
Hide thumbs Also See for SP Series:
Table of Contents

Advertisement

Figure 8
Change filter element
To start the maintenance work, please follow these steps:
SP190-H/DIL-EX2 T* with weather protection cover: Open the two clamps and remove protection
cover from the sample dilution probe.
Loosen filter lid
element sealings, filter element with filter element holder and filter's knurled head thumb screw.
Note
Unscrew the knurled head thumb screw of the filter and remove the filter element.
Check the seals of the filter element and replace if necessary.
Check O-rings of the filter lid and replace if necessary.
Place new filter element on the filter element holder.
Screw-on the knurled head thumb screw again.
Clean the filter chamber. Now, while the filter lid with the filter element is removed, you can clean
the inside of the probe tube to remove any residue.