Connections
Leader
Follower
Figure 6
Connections for synchronization and reference clock skew calibration
Connections for Synchronization
The following two connections are required for synchronization.
1
2
Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide
Connect TRIG OUT from the Leader to AUX TRIG IN on the Follower through a
BNC cable.
Connect 10 MHz REF OUT from the Leader to 10 MHz REF IN on the Follower
through a BNC cable.
Two-Oscilloscope System, MXR/EXR-Series
Front
4
Side
GEN
OUT
AUX
OUT
10 MHz
REF IN
10 MHz
REF OUT
AUX
TRIG IN
TRIG
OUT
GEN
OUT
AUX
OUT
10 MHz
REF IN
10 MHz
REF OUT
AUX
TRIG IN
TRIG
OUT
25