N5247A/B or N5227A/B RF Interface Cable Connections (U3021-60101)

N5247A/B or N5227A/B RF Interface Cable Connections (U3021-60101)
Figure 10
analyzer. The cables have been supplied with Cable Kit (U3021-60101).
1. Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM
2. Connect the RF interconnect cables from the network analyzer to the test
Over torque will cause damage to the test set and may cause connectors to
spin or become loose.
Table 7
Numeric
Order
1
2
3
4
Figure 10
The cables must be connected in the numeric order listed in
Keysight U3047AM12 User's and Service Guide
illustrates the cable configuration of the test set to the network
jumpers (x8) on the network analyzer. The RCVR R1 to SOURCE OUT
reference loop jumpers (x4) remain on the front panel.
set in the order listed. As you are connecting each cable, torque to 8 in-lb.
Refer to
Table 7
and
Figure 10
RF Interface Cable Connection (U3021-60101)
RF Cables
U3047-20016
U3047-20016
U3047-20017
U3047-20017
illustrates the final two digits of the part number for each cable.
.
From Network
To Test Set
Analyzer
CPLR ARM
CPLR ARM
RCVR IN
RCVR OUT
CPLR THRU
CPLR THRU
SOURCE OUT
SOURCE IN
Table 7
.