Connecting The Oscilloscopes - Keysight Technologies UXR Series Hardware Configuration Manual

Multiscope
Table of Contents

Advertisement

2
Two-Oscilloscope System, UXR-Series

Connecting the Oscilloscopes

Using the parts in a N2164A (two in rack) expansion kit, follow these instructions:
1
2
3
4
5
6
14
Leader
Follower #1
Power divider
Mount the 54964-60401 power divider assembly in the designated rack
location (with connectors facing back). See
Rack"
on page 12.
Connect the four SMA 50
left-most power divider assembly outputs.
Connect one 54964-61632 cable from the Leader Sample_CLK_Out output to the
power divider input.
Connect two 54964-61632 cables between the power divider assembly outputs
and the oscilloscopes' Sample_CLK_In inputs.
Connect one 54964-61632 cable between the Leader Trig Out and Follower #1
Aux Trig In connectors.
Connect one 54964-61632 cable between the Leader Reference CLK Out and
Follower #1 Reference CLK In connectors.
Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide
Connections
(Back)
Aux Trig
Trig Out
In
Sample
Reference
10 MHz
CLK
CLK
Out In
In Out
In Out
Aux Trig
Trig Out
In
Sample
Reference
10 MHz
CLK
CLK
Out In
In Out
In Out
"Mounting Oscilloscopes in
terminators (1250-3745) to the four unused
Ω
Place 1250-3745 terminators
on all unused divider outputs

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mxr seriesExr series