Keysight Technologies UXR Series Hardware Configuration Manual

Keysight Technologies UXR Series Hardware Configuration Manual

Multiscope

Advertisement

Quick Links

Keysight UXR-Series and
MXR/EXR-Series MultiScope
Hardware
Configuration
Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UXR Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Keysight Technologies UXR Series

  • Page 1 Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...
  • Page 2 12.212 and 27.405-3 and Department of prior agreement and written consent from Defense FAR Supplement ("DFARS") Keysight Technologies, Inc. as governed by CAUTION 227.7202, the U.S. government acquires United States and international copyright commercial computer software under the laws.
  • Page 3: Multiscope Hardware Configuration, Uxr-Series-At A Glance

    MultiScope Hardware Configuration, UXR-Series—At a Glance In one equipment rack, you can connect multiple Infiniium UXR-Series oscilloscopes together to create an oscilloscope system with up to 20-channels. The Infiniium UXR-Series oscilloscope software supports up to 40 channels in a MultiScope N OTE system.
  • Page 4 For example, to set up a four-oscilloscope system, your configuration would include one N2164A kit (to go from one oscilloscope to two) and two N2165A kits (to go from two oscilloscopes to three and then from three oscilloscopes to four). Figure MultiScope To control MultiScope systems and analyze captured data, you can use:...
  • Page 5: Multiscope Hardware Configuration, Mxr/Exr-Series-At A Glance

    MultiScope Hardware Configuration, MXR/EXR-Series—At a Glance You can connect two Infiniium MXR/EXR-Series oscilloscopes together to create an oscilloscope system with up to 16 channels. Oscilloscopes Infiniium MXR/EXR-Series oscilloscopes: Required • Each oscilloscope adds eight or four input channels. You can connect two oscilloscopes in a rack.
  • Page 6: In This Guide

    In This Guide This guide describes the equipment required to configure MultiScope systems and shows you how to set up and connect the equipment: • Chapter 1, “MultiScope System Requirements,” starting on page Depending on the UXR-Series oscilloscope bandwidths and number of channels required, you will use one of the configurations shown in the following table.
  • Page 7: Table Of Contents

    Contents MultiScope Hardware Configuration, UXR-Series—At a Glance / 3 MultiScope Hardware Configuration, MXR/EXR-Series—At a Glance / 5 In This Guide / 6 MultiScope System Requirements Two-Oscilloscope System, UXR-Series Expansion Equipment Required / 11 Mounting Oscilloscopes in Rack / 12 Connecting the Equipment / 13 Parts in the N2164A (Two in Rack) Expansion Kit / 13 Connecting the Oscilloscopes / 14 Three- to Five-Oscilloscope Systems, UXR-Series...
  • Page 8 Infiniium Oscilloscope Software on the Leader Oscilloscope / 30 D9010BSEO Infiniium Offline Oscilloscope Analysis Software / 30 Go To “Using Hosted Infiniium Oscilloscopes” in the Online Help / 31 Index Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...
  • Page 9: Multiscope System Requirements

    Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide 1 MultiScope System Requirements This chapter describes the requirements for MultiScope systems, including oscilloscopes, environmental, and rack. Oscilloscopes are sensitive measurement instruments, and some environmental conditions must be present to ensure that the oscilloscope delivers the best possible measurement accuracy.
  • Page 10 MultiScope System Requirements Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...
  • Page 11: Two-Oscilloscope System, Uxr-Series

    Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide 2 Two-Oscilloscope System, UXR-Series Expansion Equipment Required / 11 Mounting Oscilloscopes in Rack / 12 Connecting the Equipment / 13 Expansion Equipment Required Table 1 Equipment Required for Two-Oscilloscope Rack Mount System Model Description Quantity...
  • Page 12: Mounting Oscilloscopes In Rack

    Two-Oscilloscope System, UXR-Series Mounting Oscilloscopes in Rack Oscilloscopes take 7U of rack space and the power divider assembly takes 1U of rack space. Locate oscilloscopes in the rack so that the overall rack center of gravity is about half-way up N OTE the rack.
  • Page 13: Connecting The Equipment

    Two-Oscilloscope System, UXR-Series Connecting the Equipment Parts in the N2164A (Two in Rack) Expansion Kit In the MBASE expansion kit, these parts are used to connect the oscilloscopes in the MultiScope system: Part number Description 54964-61632 1 m phase-stable cable 54964-60401 6-way splitter assembly with rack mount shelf 50 Ω...
  • Page 14: Connecting The Oscilloscopes

    Two-Oscilloscope System, UXR-Series Connecting the Oscilloscopes Connections (Back) Aux Trig Trig Out Leader Sample Reference 10 MHz Out In In Out In Out Aux Trig Trig Out Follower #1 Sample Reference 10 MHz Out In In Out In Out Place 1250-3745 terminators Power divider on all unused divider outputs Using the parts in a N2164A (two in rack) expansion kit, follow these instructions:...
  • Page 15 Two-Oscilloscope System, UXR-Series Connect the oscilloscopes to the Leader oscilloscope (or external host computer) through the desired remote interface connection (LAN, USB, etc.). If these are the last oscilloscopes to be connected, see Chapter 5, “Next Steps,” starting on page 27. Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...
  • Page 16 Two-Oscilloscope System, UXR-Series Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...
  • Page 17: Three- To Five-Oscilloscope Systems, Uxr-Series

    Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide 3 Three- to Five-Oscilloscope Systems, UXR-Series Expansion Equipment Required / 18 Mounting Oscilloscopes in Rack / 19 Connecting the Equipment / 20...
  • Page 18: Expansion Equipment Required

    Three- to Five-Oscilloscope Systems, UXR-Series Expansion Equipment Required Table 2 Equipment Required for 3-to-5-Oscilloscope Rack Mount System Model Description Quantity N2164A UXR-Series Multi-Frame Base Kit (First Two Frames) N2165A UXR-Series Multi-Frame Expansion Kit (Add One (# of scopes) - 2 Frame) N2156A Rack mount kit...
  • Page 19: Mounting Oscilloscopes In Rack

    Three- to Five-Oscilloscope Systems, UXR-Series Mounting Oscilloscopes in Rack Oscilloscopes take 7U of rack space and the power divider assembly takes 1U of rack space. Locate oscilloscopes in the rack so that the overall rack center of gravity is about half-way up N OTE the rack.
  • Page 20: Connecting The Equipment

    Three- to Five-Oscilloscope Systems, UXR-Series Connecting the Equipment Parts in the N2165A (Add One) Expansion Kit Part number Description 54964-61632 1 m phase-stable cable Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...
  • Page 21: Connecting The Follower #2 Through Follower #4 Oscilloscopes

    Three- to Five-Oscilloscope Systems, UXR-Series Connecting the Follower #2 Through Follower #4 Oscilloscopes Connections (Back) Aux Trig Trig Out Leader Sample Reference 10 MHz Out In In Out In Out Aux Trig Trig Out Follower #1 Sample Reference 10 MHz Out In In Out In Out...
  • Page 22 Three- to Five-Oscilloscope Systems, UXR-Series Before connecting the follower #2 through follower #4 oscilloscopes, first follow the instructions for connecting a two-oscilloscope rack mount system in “Connecting the Oscilloscopes" on page 14. Then, for each of the oscilloscopes from follower #2 to follower #4, using the parts in a N2165A (add one) expansion kit, follow these instructions: Remove the SMA 50 terminator (1250-3745) from the power divider output...
  • Page 23: Two-Oscilloscope System, Mxr/Exr-Series

    Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide 4 Two-Oscilloscope System, MXR/EXR-Series MXR/EXR-Series Oscilloscope Requirements / 24 Connections / 25 Connections for Reference Clock Skew Calibration / 26 This chapter describes how to configure two-oscilloscope MultiScope systems made up of MXR/EXR-Series oscilloscopes.
  • Page 24: Mxr/Exr-Series Oscilloscope Requirements

    Two-Oscilloscope System, MXR/EXR-Series MXR/EXR-Series Oscilloscope Requirements For two-oscilloscope MultiScope systems made up of MXR/EXR-Series oscilloscopes, you can use the N2124A multi-frame cable kit. Refer to your analysis software documentation for additional Infiniium application software version requirements. Parts in the N2124A Multi-Frame Cable Kit 8120-1840 (qty 3), BNC cable, 48 in.
  • Page 25: Connections

    Two-Oscilloscope System, MXR/EXR-Series Connections Front Side 10 MHz REF IN 10 MHz REF OUT TRIG IN Leader TRIG 10 MHz REF IN 10 MHz REF OUT TRIG IN Follower TRIG Figure 6 Connections for synchronization and reference clock skew calibration Connections for Synchronization The following two connections are required for synchronization.
  • Page 26: Connections For Reference Clock Skew Calibration

    Two-Oscilloscope System, MXR/EXR-Series From the N2124A multi-frame cable kit, use these cables: Connection Use Cable/Adapters Trig 8120-1840 BNC cable, 48 in. length Ref Clk 8120-1840 BNC cable, 48 in. length Connections for Reference Clock Skew Calibration The following connections are required for reference clock skew calibration. These same connections are also required for the optional drift correction.
  • Page 27: Next Steps

    Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide 5 Next Steps Individually Calibrate the Oscilloscopes in the MultiScope System / 27 UXR-Series MultiScope Calibration Connections / 28 Use MultiScope Control Software / 30 Go To “Using Hosted Infiniium Oscilloscopes” in the Online Help / 31 With UXR-Series MultiScope systems, you make a long calibration cable and later follow prompts by the Infiniium oscilloscope software to connect the cable to individual oscilloscopes in the system as necessary.
  • Page 28: Uxr-Series Multiscope Calibration Connections

    Next Steps UXR-Series MultiScope Calibration Connections Part of what the MultiScope control software provides is the ability to automatically deskew multiple oscilloscope frames and channels using a calibration procedure. The hardware for performing MultiScope calibration is included in the N2164A UXR-Series Multi-Frame Base Kit.
  • Page 29 Next Steps Calibration Cable Connections (Follower #1 Front) Cal Out 1250-3377 push-on SMA adapter You will connect this cable-end to Leader and Follower oscilloscope 1250-2818 inputs as prompted during the right-angle “Calibrate” procedure in the Infiniium SMA adapter software’s Hosted Setup dialog box. 54964-61632 cable 1250-2818...
  • Page 30: Use Multiscope Control Software

    Next Steps Use MultiScope Control Software After the MultiScope system hardware has been configured and you have performed each oscilloscope’s calibration procedure, you can use MultiScope control software to make data acquisitions. Your options are: • Infiniium oscilloscope software on the Leader oscilloscope. MultiScope software is included as a standard feature.
  • Page 31 Next Steps Go To “Using Hosted Infiniium Oscilloscopes” in the Online Help For instructions on: • Identifying oscilloscopes in the MultiScope system (making software connections). • Time-correlating oscilloscopes (calibration). • Using the MultiScope system. Go to the “Using Hosted Infiniium Oscilloscopes” topic in the Infiniium oscilloscope software’s Online Help.
  • Page 32 Next Steps Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...
  • Page 33 Index two UXR-Series oscilloscope rack mount system, at a glance, MultiScope configuration, 3, in the box, MultiScope control software, IO Libraries Suite, Cal Out signal, 3, calibrate oscilloscopes in system, MultiScope hardware configuration, 3, computer (host), 4, MultiScope, product overview, 3, computer (host), requirements, MXR/EXR-Series oscilloscopes, Connection Expert,...
  • Page 34 Index Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide...

This manual is also suitable for:

Mxr seriesExr series

Table of Contents