Three- to Five-Oscilloscope Systems, UXR-Series
Before connecting the follower #2 through follower #4 oscilloscopes, first follow
the instructions for connecting a two-oscilloscope rack mount system in
"Connecting the Oscilloscopes"
Then, for each of the oscilloscopes from follower #2 to follower #4, using the parts
in a N2165A (add one) expansion kit, follow these instructions:
1
2
3
4
5
6
22
Remove the SMA 50
Ω
that will be used for the oscilloscope.
Connect one 54964-61632 cable between the power divider output and the
oscilloscope's Sample_CLK_In input.
Connect one 54964-61632 cable between the previous oscilloscope's Trig Out
connector to this oscilloscope's Aux Trig In connector.
Connect one 54964-61632 cable between the previous oscilloscope's
Reference CLK Out connector to this oscilloscope's Reference CLK In connector.
Connect the oscilloscope to the Leader oscilloscope (or external host
computer) through the desired remote interface connection (LAN, USB, etc.).
If this is the last oscilloscope to be connected, see
starting on page 27.
Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide
on page 14.
terminator (1250-3745) from the power divider output
Chapter
5, "Next Steps,"