4.4.2 Set the Internal Resistance
This series instrument provides internal resistance setting (CV priority mode on-
ly). The procedures are shown as below.
1. Press the composite keys [Shift]+[V-set] (Config) on the front panel to enter
2. Rotate the knob to select Output Res and press [Enter] to confirm.
3. Set the internal resistance value and press [Enter] to confirm.
4.4.3 CR Function in Sink Mode
This series of instruments supports CR testing in sink mode, and the sink cur-
rent capability of CR is controlled by the Is setting value in CC priority mode.
The usage of this function is as follows:
1. Press the composite keys [Shift]+[V-set] (Config) on the front panel to enter
2. Select Mode and press [Enter].
3. Select CC and press [Enter].
4. Select High or Low, and press [Enter].
5. Rotate the knob to set the current rise and fall time.
6. Rotate the knob to select Sink Res and press [Enter].
7. Select On and press [Enter].
8. Set the Sink Res value and press [Enter].
9. Set the voltage and current.
the configuration menu.
the configuration menu.
Enter the setting interface of the loop response speed.
At this point, return to the Config menu settings interface.
If set to 0, the function is equivalent to that Sink Res is selected as OFF.
The following takes the setting of 10Ω as an example to introduce the actual
test results.
a. Press the [I-set] key on the front panel to set the output current value Is.
Take –5A as an example.
b. Press the [V-set] key on the front panel to set the voltage upper limit Vh=
20V, and press [V-set] again to set the voltage lower limit Vl=0V.
c. Set the DUT (power supply) to output 80V, 10A.
At this time, 10Ω in sink mode (according to the formula I=U/R, sink current
is 80÷10=8A), due to the limitation of Is=–5A in CC priority mode, the actual
Copyright © Itech Electronic Co., Ltd.
Power Supply Function
52
Need help?
Do you have a question about the IT-M3900C Series and is the answer not in the manual?
Questions and answers