Model 512 Instruction Manual

TEST

In Test mode, one complete scan is performed on the circuit under test.
configuration compares exactly with the "learned" continuity program in memory, the test is
successful and the display will show "Passed." If an error is encountered, the points in error are
displayed, along with an indication of OPEN or SHORT. If a printer is connected, and it is
desirable to record errors, use ERROR LIST instead of, or after, this procedure. A complete Test
procedure is provided below. Once familiarized, however, the user may refer to Figure 5. To
select and use TEST, proceed as follows:
1. Load program to be used (refer to "Load Program").
2. Install harness to be tested.
3. Press TEST key; Single Test will be displayed, and one complete scan will be made.
4. If Passed appears at the end of the scan, the test is successful.
5. An OPEN or SHORT, with the From/To points, will be displayed when an error is
encountered.
6. To continue test beyond the failed points, press TEST again and testing will resume.
7. When testing is finished, disconnect harness.

RESET

The RESET mode is available for terminating any mode currently in operation. Pressing RESET
is required with some modes that display an end message upon completion. When RESET is
pressed, the display will read Cablescan Model 512 Ready.