Press the M Mode key to acquire a B/M- Mode image.
Press CONTROL Q to go to B/A Mode. The B -Mode
M
image appears on the left side and the A-Mode image
appears on the right side. To exit the B/A Mode, press
CONTROL Q again and it reverts to B/M Mode. Press
the B Mode key to return to the original display. Not
exiting the B/A Mode by pressing CONTROL Q causes
the system to default back to B/A Mode when the
M–Mode key is pressed in B–Mode and M–Mode.
.
Illustration 15. B/A Mode Display
NOTE: When the B Mode key is pressed in B/A Mode,
the system returns to B-Mode retaining current scan
values. It erases measurements. If M Mode key is
pressed now the system goes back to B/A Mode.
When M Mode key is pressed in B/A Mode, the system
reverts to M Mode and if M Mode key is pressed again it
goes back to B/A Mode.
Only Distance Measurements are possible in B/A Mode.
A/B Ratio is the only calculation that is allowed. Invoking
Heart Rate, Velocity or Volume measurements causes
the system to beep and displays and ILG message.
Dynamic Range, Focus, Depth and Gain Control
adjustments affect the B/A-Mode image.