Distance Measurement - GE LOGIQ A100 Technical Publication

Table of Contents

Advertisement

Distance Measurement

Distance Measurements are made in the B-Mode portion
of the image. To make a distance measurement:
Press FREEZE to stop image acquisition. (It is not
A "
mandatory while making a distance measurement).
Press the MEASUREMENT key once to display the
Meas
General Measurement Menu. Simultaneously an open
ended " " cursor appears on the screen.
Use the TRACKBALL to move the open ended " " cursor
to the measurement start point.
TB
Press SET to fix the measurement start point cursor and
to display a second cursor. Distance in mm appears on
the left side of the screen.
Use the TRACKBALL to move the second open ended
TB
" " cursor to the measurement end point.
Press SET to complete the measurement. The Distance
value is displayed in mm on the left of the screen.
Repeat the same procedure to make more Distance
Measurements, if required. Up to 4 measurements will
be displayed in mm at a time .
To enable the already fixed cursor to become active,
Hints
press the MEASUREMENT key; the fixed cursor
becomes active and the currently active cursor becomes
fixed.
t
α 100 User Manual
LOGIQ
2211157–100 Rev 0